• 中文核心期刊
  • 中国科技核心期刊
  • CSCD来源期刊
  • DOAJ 收录
  • Scopus 收录

微波消解-高分辨电感耦合等离子体质谱法测定土壤中8种金属元素

Determination of 8 Metal Elements in Soil by High-resolution Inductively Coupled Plasma-Mass Spectrometry with Microwave Digestion

  • 摘要: 微波消解结合电感耦合等离子体质谱(ICP-MS)是土壤样品中金属元素测定的常用方法,其前处理可以采用不同的消解体系,但是消解体系对分析结果的准确性影响较大。此外,应用ICP-MS测定某些元素时干扰的存在会影响结果的准确性。基于上述问题,本文优选三个酸体系微波消解溶样,采用高分辨电感耦合等离子体质谱(HR-ICP-MS)测定土壤样品中8种金属元素(Cr、Co、Ni、Cu、Zn、Pb、Cd和U)的含量,对比研究了由不同用量硝酸、盐酸、氢氟酸混合组成的三个酸消解体系对国家土壤标准物质的消解效果,确定了最优前处理方法。结果表明:经国家土壤标准物质验证,采用HR-ICP-MS检测,在不需要干扰校正的情况下,酸体系Ⅰ(6mL硝酸+3mL盐酸+3mL氢氟酸)和酸体系Ⅱ(2mL硝酸+6mL盐酸+1mL氢氟酸)的测定值与标准值相吻合,方法检出限为0.001~0.715μg/g,精密度(RSD,n=6)小于7.0%。从消解情况、准确度和精密度比较,酸体系Ⅰ稍优于酸体系Ⅱ;从酸用量比较,酸体系Ⅱ酸用量最少。两种酸体系的样品处理方法均具有较高的适用性和可靠性,都可用于土壤样品中8种金属元素含量的直接测定。

     

    Abstract:
    BACKGROUND  Microwave digestion combined with inductively coupled plasma-mass spectrometry (ICP-MS) is a commonly used method for the determination of metal elements in soil samples. Different solvents can be used in the pretreatment, and the digestion method has a great influence on the accuracy of the analysis results. In addition, the accuracy of the results will be affected by the interference in the determination of some elements by ICP-MS.
    OBJECTIVES  To accurately determine metal elements in soil samples by high resolution-inductively coupled plasma-mass spectrometry (HR-ICP-MS).
    METHODS  Soil samples were digested by microwave using three different solvents, and 8 metal elements (Cr, Co, Ni, Cu, Zn, Pb, Cd and U) were determined by HR-ICP-MS. The digestion effects of the three acid digestion methods using different amounts of nitric acid, hydrochloric acid and hydrofluoric acid on certified soil reference materials were studied, and the optimal pretreatment procedure was determined.
    RESULTS  The proposed procedures have been verified by national soil reference materials. It was found that the measured values of digestion method Ⅰ(6mL HNO3+3mL HCl+3mL HF) and digestion method Ⅱ(2mL HNO3+6mL HCl+1mL HF) were consistent with the certified values, using HR-ICP-MS without interference correction. The detection limits of both procedures were 0.001-0.715μg/g. The relative standard deviations (RSD, n=6) were all less than 7.0%. In terms of digestion, accuracy and precision, acid system Ⅰ was slightly better than acid system Ⅱ, but acid system Ⅱ had the least amount of acid.
    CONCLUSIONS  The two preferred procedures have high applicability and reliability, and can be used for the direct determination of 8 metal elements in soil samples.

     

/

返回文章
返回