Feng WEI, Jing-ke HU. Study on Residues and Cleaning of Organochlorine Pesticides and Polychlorinated Biphenyls in Microsyringes[J]. Rock and Mineral Analysis, 2014, 33(4): 578-583.
Citation: Feng WEI, Jing-ke HU. Study on Residues and Cleaning of Organochlorine Pesticides and Polychlorinated Biphenyls in Microsyringes[J]. Rock and Mineral Analysis, 2014, 33(4): 578-583.

Study on Residues and Cleaning of Organochlorine Pesticides and Polychlorinated Biphenyls in Microsyringes

More Information
  • Received Date: July 18, 2013
  • Revised Date: September 15, 2013
  • Accepted Date: April 08, 2014
  • Published Date: June 30, 2014
  • Microsyringes are being increasingly used to transfer trace liquid, but the concern of cross contamination caused by residues of organic contaminant in microsyringes is routinely overlooked. The residues and cleaning effects of organochlorine pesticides (OCPs) and polychlorinated biphenyls (PCBs) in microsyringes were systematically researched and the findings are presented in this paper. Results show that the cleaning effects are different for three brands of microsyringe. Microsyringes should be chosen according to the suitability to the intended experiment. The volume of cleaning solvent was proportional to the dilution ratio of organic contaminant, which was due to the approximate amount of contaminant in the first cleaning fluid by 4 volumes (0.1 mL, 0.2 mL, 0.5 mL and 1 mL). Moreover, the volume of cleaning solvent had a negative correlation with the amount of washout compounds. The best effect and lowest cost came from using 0.1 mL cleaning solvent. There is no difference in cleaning effects of OCPs and PCBs standard solution at the same concentration probably due to the similar physical and chemical properties of the two. Three cleaning observations were made: the residue amount of OCPs or PCBs in a 1 mL microsyringe was about 1% of the loaded amount, the amount in the first cleaning solvent was over 90% of the residues, and with the increase in cleaning times, the cleaning effect was less effective, and dilution ratio smaller. According to the above, corresponding cleaning methods can be used, according to varying concentrations of different pollutants, as far as possible to reduce organic contaminant in microsyringes and therefore reduce the cost.
  • Han D D, Kyung H R.Trends in liquid-phase microextraction, and its application to environmental and biological samples[J].Microchim Acta, 2012, 176:1-22. doi: 10.1007/s00604-011-0678-0
    Shahab S F, Yadollah Y, Naader B. Headspace solvent microextraction and gas chromatographic determination of some polycyclic aromatic hydrocarbons in water samples[J].Analytica Chimica Acta, 2003, 489:21-31. doi: 10.1016/S0003-2670(03)00709-8
    王素利,杨素萍,刘丰茂,薛佳莹,尤祥伟.液相微萃取技术在农药残留分析中的应用研究进展[J].农药学学报, 2012, 14(5):461-474. http://www.cnki.com.cn/Article/CJFDTOTAL-NYXB201205002.htm
    路雪芹,梁红霞.加药注射器使用方法与细菌污染相关性实验研究[J].护理研究, 2004, 18(1):157. http://www.cnki.com.cn/Article/CJFDTOTAL-SXHZ200402050.htm
    夏余芝.医用注射器结构不合理性的分析[J].疾病控制杂志, 2004, 8(5):450-451. http://www.cnki.com.cn/Article/CJFDTOTAL-JBKZ200405024.htm
    王希平.关于安全注射器安全性能的研究——兼论"安全注射器的评定标准"[J].中国医疗器械信息, 2006, 12(11):46-65. doi: 10.3969/j.issn.1006-6586.2006.11.016
    程新民.微量注射器的正确使用及对色谱分析的影响[J].石油化工, 1990, 19(9):636-638. http://www.cnki.com.cn/Article/CJFDTOTAL-SYHG199009010.htm
    郭晓君,曹占文,李建忠.粮油检验的气相色谱法中微量注射器的使用方法[J].吉林粮食高等专科学校学报, 2001, 16(2):1-3. http://www.cnki.com.cn/Article/CJFDTOTAL-JLZX200102000.htm

Catalog

    Article views (779) PDF downloads (10) Cited by()

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return