石英缝管技术火焰原子吸收测定化探样品中痕量金
Determination of Gold by Slotted Quartz Tube AAS
-
摘要: 本文研究了应用石英缝管技术火焰原子吸收测定化探样品中痕量Au的方法。考查了不同规格带缝石英管以及各种不同条件对测定灵敏度的影响。实验表明,当应用5cm缝长燃烧器、石英缝管缝长为6cm左右、石英管狭缝与燃烧器狭缝的距离为3mm时,得出的灵敏度最高。应用石英缝管技术测定Au可提高灵敏度2.5倍左右。Abstract: A method for determination of gold by flame-AAS with slotted quartz tube has been developed. The effect of various conditions on the sensitivity of gold were investigated. The best sensitivity was obtained with a burner slot of 5cm long and a quartz slot
-
Keywords:
- gold /
- determination /
- slotted /
- quartz /
- tube
-
计量
- 文章访问数: 2012
- HTML全文浏览量: 0
- PDF下载量: 1025