Preparation and Certification of A Titanium Isotopic Standard Solution
-
摘要: 钛同位素组成可用于地球化学和宇宙化学研究中,但目前国内外缺乏钛同位素标准物质。为了满足地质样品钛同位素分析的需要,本文研制了钛同位素标准溶液,报道了钛同位素标准溶液的研制过程和定值结果,包括标准溶液的选择、均匀性和稳定性检验、定值分析及测定数据的统计性检验等。通过初步测定,确定美国Alfa公司生产的Ti单元素溶液作为备选Ti同位素标准溶液。将备选标准溶液分装成150瓶,随机抽取15瓶进行均匀性检验,测试结果的F值均小于临界值,表明备选标准溶液的Ti同位素组成均匀。通过30个月的稳定性检验,标准溶液的特征量值变化在不确定度范围内。采用独家和多家实验室相结合的方法进行定值,标准溶液的特征量值及不确定度推荐为:δ50Ti=-2.23‰±0.14‰,δ49Ti=-1.67‰±0.09‰,δ48Ti=-1.13‰±0.06‰,δ47Ti=-0.57‰±0.05‰。研制的标准溶液可用于钛同位素分析时校正仪器和验证质谱分析过程,有利于不同实验室的测试数据之间的对比和应用。Abstract: Titanium isotopic compositions have been used to elucidate geochemical and cosmochemical processes. Until now, no Ti isotopic reference material has been reported worldwide, which imposes restriction upon analytical method establishment and laboratory quality control. In this paper, a newly prepared Ti isotopic standard solution is introduced. Ti element standard solution offered by the Alfa Company from U.S.A. was chosen as the starting material, based on the comparison of Ti isotopic measurement results of 3 commercially available Ti element standard solutions. The preparation of the new standard solution was strictly followed using the stipulation on reference material of the metrological technical standard of state, including homogeneity and stability tests, and certification analyses. 150 bottles of the solution were prepared, among which 15 bottles were selected randomly for homogeneity test and no detectable heterogeneity was found by F test. The stability inspection through 30 months indicated no significant changes on δTi values. The certified values and uncertainty at the 95% confidence level were shown to be: δ50Ti=-2.23‰±0.14‰, δ49Ti=-1.67‰±0.09‰,δ48Ti=-1.13‰±0.06‰ andδ47Ti=-0.57‰±0.05‰. This newly certified standard solution can be used to calibrate instruments and verify analytical quality in Ti isotope measurements, and compare inter-laboratory data.
-
Keywords:
- titanium isotopes /
- standard solution /
- certified value
-
电感耦合等离子体质谱(ICP-MS)具有灵敏度高、干扰少、多元素同时测定、线性范围大、检出限低的特点,适合地质样品中微量级多元素同时测定。对于微量元素的测定通常采用氢氟酸、硝酸在封闭溶样罐中高温、高压分解,该方法能有效分解岩石、矿物中的难溶矿物[1-2],由于ICP-MS仪器的高灵敏度,大部分微量元素的检出限可达到化探样品分析要求。Ag有两个同位素,107Ag(51.84%)和109Ag(48.16%),它们分别受到91Zr16O和93Nb16O氧化物离子的强烈干扰,由于化探样品中Zr和Nb含量大大高于Ag,即使用Zr和Nb的单标准氧化物产率进行校正,其结果误差仍然较大,因为其干扰信号强度已经超过了样品中Ag本身的强度。虽然使用膜去溶装置可以大大降低氧化物离子干扰,实现Ag的准确测定[3];但该装置价格高,拥有膜去溶装置的实验室较少。目前对于化探样品Ag的测定,国内大部分实验室仍然采用传统的发射光谱法[4-9],该方法费时、费力、结果不稳定。因此,迫切需要对化探样品Ag的测定方法进行改进。
本文应用P507萃淋树脂将用于ICP-MS测定常规微量元素的溶液进行简单的分离,干扰元素Zr和Nb可被有效除去,而Ag和内标元素Rh可被有效回收,实现了化探样品中低含量Ag的快速测定。
1. 实验部分
1.1 仪器及工作条件
Bruker Aurora M90电感耦合等离子体质谱仪(美国布鲁克·道尔顿公司)。在每次测试前,为了降低仪器本底,样品锥和截取锥都需仔细清洗。在5%硝酸溶液中Ag的仪器背景值通常都小于50 cps,使用普通灵敏度模式,仪器灵敏度通常调整为>400000 cps/1 ng/mL 115In,相对标准偏差(RSD)通常小于3%。本实验采用雾化器自吸进样,仪器工作参数见表 1。
1.2 材料与主要试剂
P507萃淋树脂:粒径80 ~120目(北京瑞乐康分离科技有限公司)。
表 1 仪器工作参数Table 1. nstrumental operating parameters of ICP-MSI工作参数 设定值 射频功率 1400 W 反射功率 < 2 W 等离子体气 15 L/min 辅助气 0.90 L/min 护鞘气 0.25 L/min 雾化气 0.95 L/min 扫描次数 5 测定次数 5 每个质量通道数 1 测定方式 Peak Hopping 停留时间 10 ms 样品锥孔径 1 mm 截取锥孔径 0.4 mm 雾化室温度 3℃ 交换柱:采用5 mL塑料移液枪头作为交换柱,底部垫自制聚四氟乙烯棉;称取0.45 g的P507萃淋树脂于烧杯中,加入约5 mL水,转移至交换柱中,待水流尽时,在上部垫一层自制聚四氟乙烯棉;用10 mL的3 mol/L硝酸淋洗,最后用5 mL的5%硝酸平衡交换柱,待用。
封闭溶样器:自制不锈钢-聚四氟乙烯封闭溶样装置,体积10 mL[1]。
多元素混合标准储备溶液:100 μg/mL (Accu-Standard Inc,USA)。
硝酸:通过石英亚沸蒸馏提纯。
氢氟酸:采用聚四氟乙烯对口瓶亚沸蒸馏提纯,实验用水用Millipore纯化装置制备,电阻率18 MΩ·cm。
1.3 实验步骤
准确称取0.0500 g样品于带不锈钢外套的聚四氟乙烯密封溶样装置中,加入1 mL氢氟酸和1 mL硝酸,加盖密封,在烘箱中于185℃加热12 h,取出冷却后在电热板上低温蒸干。最后加入2 mL硝酸、1 mL 500 ng/mL的Rh内标溶液、3 mL水,重新盖上盖密封,放入烘箱中于135℃加热3 h溶解残渣。冷却后取0.4 mL溶液于15 mL离心管中,用5%硝酸稀释至6 mL。该溶液可用于ICP-MS测定常规微量元素。
待微量元素测定完成后,将剩余溶液倒入交换柱中,直至加满交换柱,其余溶液弃去,并立即用水清洗离心管,用原离心管承接,该溶液即可用于以Rh为内标Ag的测定。
2. 结果与讨论
2.1 Ag与Zr和Nb的分离
P507是酸性磷类萃取剂,又名2-乙基己基膦酸单2-乙基己基酯,常用于稀土元素分离以及稀土元素的相互分离[10-13],在Sm-Nd同位素测定中也常用P507或P204萃淋树脂实现Sm与Nd的相互分离[14]。该树脂的另一个特点是对Ti、Nb、Ta、Zr、Hf、W、Sn和Mo等元素的四价离子强烈吸附,即使用高浓度的盐酸或硝酸也很难将其洗脱下来,只有用氢氟酸才能将这些元素有效洗脱,该类树脂也可用于Lu-Hf同位素分离[15-16]。本研究利用该树脂这一特性,在约1.2 mol/L的硝酸介质中成功地实现了Ag和内标元素Rh与干扰元素Zr和Nb的有效分离。
取200 ng混合标准溶液于15 mL离心管中,用1.2 mol/L硝酸稀释至5 mL,将此溶液过柱,15 mL离心管承接,用4 mL的5%硝酸分两次清洗离心管及交换柱,在承接溶液的离心管中加入100 ng的Rh 内标溶液,最后稀释至10 mL,ICP-MS测定。各元素的回收率见表 2。由表 2可以看出,98%以上的Zr和Nb被P507树脂吸附,而95%以上的Ag和Rh通过交换柱,说明P507萃淋树脂能有效地将Ag和Rh与Zr和Nb分离。
表 2 各元素在P507萃淋树脂上的回收率Table 2. The recovery of elements for P507 levextrel resin元素 回收率/% Zr 0.93 Nb 1.44 Mo 2.83 Sn 0.43 Hf 0.58 Ta 0.42 W 3.32 Cd 103.0 Ag 95.8 Rh 97.5 2.2 方法检出限
按样品前处理同样程序处理5份流程空白,测定结果见表 3。其绝对浓度值3倍标准偏差除以称样量,即为方法的检出限,计算Ag的检出限为0.005 μg/g,低于化探样品分析的检出限要求(0.02 μg/g,见DZ/T 0130.5—2006)。
表 3 方法的空白值Table 3. Blank level of the method空白 m(Ag)/μg 空白1 0.0004 空白2 0.0002 空白3 0.0003 空白4 0.0003 空白 m(Ag)/μg 空白5 0.0004 平均值 0.0003 标准偏差 0.000075 2.3 树脂的再生
交换柱使用后立即用水洗柱一次,然后用3 mol/L硝酸5 mL洗柱1次,再用水洗柱两次,最后用1.2 mol/L硝酸5 mL平衡交换柱,待用。树脂在使用一段时间后,其吸附的Ti、Zr、Nb等元素可能达到饱和,这时树脂吸附效率会降低。一般在使用5~10次后,用2 mol/L氢氟酸5 mL将这些元素洗脱下来,树脂即可继续使用。如果发现用氢氟酸洗脱后交换柱的效率仍然很低,说明P507萃取剂已流失,这时需要更换新树脂。
3. 标准物质分析
按上述分析流程对岩石及土壤系列国家一级标准物质进行分析,本方法的测定结果与标准值基本一致(见表 4),完全能够满足化探样品分析要求。
表 4 标准物质测定结果Table 4. Analytical results of Ag in reference materials标准物质
编号w(Ag)/(μg·g-1) 标准值 本法测量值 GBW 07103 0.033±0.010 0.026±0.008 GBW 07104 0.071±0.014 0.065±0.010 GBW 07105 0.040±0.012 0.051±0.009 GBW 07106 0.062±0.010 0.055±0.007 GBW 07302 0.066±0.015 0.065±0.008 GBW 07305 0.36±0.04 0.35±0.02 GBW 07306 0.36±0.04 0.31±0.05 GBW 07307 1.05±0.09 1.12±0.08 GBW 07311 3.2±0.5 3.02±0.32 GBW 07312 1.15±0.16 0.99±0.11 4. 结语
利用ICP-MS仪器的碰撞池技术可以消除氧化物离子干扰,但仪器灵敏度会降低,碰撞气体有可能带入新的干扰,Ag也需要进行单独测定;膜去溶装置可去除气溶胶中的大部分水分,降低氧化物离子干扰,提高仪器灵敏度,实现Ag与其他常规微量元素同时测定,但进样时间可能延长,设备也较贵。而本文应用P507萃淋树脂对ICP-MS用于测定常规微量元素的溶液进行简单分离,就可实现化探样品中待测元素Ag和内标元素Rh与干扰元素Zr、Nb的有效分离,Ag的检出限达到0.005 μg/g,低于化探分析要求(0.02 μg/g)。
相比于其他方法,本方法省略了称样及分解等样品前处理步骤;且由于在样品处理过程中加入了内标元素,因此最后的溶液不需要准确定容,待测元素与内标元素都具有很高的回收率,过柱分离的溶液只需3~4 mL即可,节省了时间,提高了分析效率。不足在之处在于:虽然本方法相对于传统的发射光谱法更简单、快速,但Ag也需要进行分离并单独测定。在本方法拓展应用方面,由于高含量W和Mo样品中W可能以单矿物形式存在,需要进行碱熔才能保证分解完全,利用P507萃淋树脂的这一特性,有可能实现W和Mo与大量基体元素和干扰元素的分离富集。
致谢: 向参加定值的厦门大学近海海洋环境科学国家重点实验室王德利研究员和张凡工程师、国家海洋局第一海洋研究所海洋环境测试中心刘季花研究员和张俊工程师表示衷心感谢。真诚感谢中国地质科学院地质研究所朱祥坤研究员和金彪工程师提供的支持和帮助。 -
表 1 三种溶液钛同位素组成初步测定结果
Table 1 Preliminary results of Ti isotope composition for three different solutions
Ti单元素溶液 δ50Ti δ49Ti δ48Ti δ47Ti Ti-1
(北京有色金属研究总院)-2.11 -1.59 -1.06 -0.54 Ti-2
(美国Alfa公司)-2.24 -1.69 -1.13 -0.55 Ti-3
(美国High-Purity Standards公司)-1.44 -1.10 -0.74 -0.37 表 2 均匀性检验方差分析结果
Table 2 Variance analysis for homogeneity inspection of reference sample
δ Q2 ν2 s22 Q1 ν1 s12 F Fα(ν1,ν2) δ50Ti 0.118 30 0.00392 0.069 14 0.00496 1.26 2.02 δ49Ti 0.070 0.00234 0.035 0.00253 1.08 δ48Ti 0.044 0.00145 0.018 0.00130 0.89 δ47Ti 0.041 0.00135 0.012 0.00082 0.61 表 3 标准溶液在30个月内特性量值的测定数据
Table 3 Measurement data of δTi in 30 months
时长/月 δ50Ti/‰ δ49Ti/‰ δ48Ti/‰ δ47Ti/‰ 4 -2.20 -1.67 -1.11 -0.54 8 -2.24 -1.69 -1.13 -0.57 12 -2.18 -1.65 -1.11 -0.56 20 -2.19 -1.66 -1.11 -0.56 22 -2.21 -1.67 -1.12 -0.56 30 -2.19 -1.65 -1.11 -0.56 表 4 标准溶液稳定性检验结果
Table 4 The stability test of reference sample
特征量值 b1 b0 s s(b1) t0.95,4×s(b1) st δ50Ti 0.000949 -2.22 0.02293 0.001056 0.00293 0.032 δ49Ti 0.000777 -1.68 0.01297 0.000597 0.00166 0.018 δ48Ti 0.000299 -1.12 0.00946 0.000436 0.00121 0.013 δ47Ti 0.000307 -0.56 0.01424 0.000655 0.00182 0.020 表 5 钛同位素标准溶液的多次测量数据(单一实验室)
Table 5 Multiple measurement data of Ti isotope composition for reference samples (unique laboratory)
序号 编号 第1次测定 第2次测定 第3次测定 平均值 δ50Ti δ49Ti δ48Ti δ47Ti δ50Ti δ49Ti δ48Ti δ47Ti δ50Ti δ49Ti δ48Ti δ47Ti δ50Ti δ49Ti δ48Ti δ47Ti 1 2 -2.14 -1.66 -1.11 -0.59 -2.15 -1.62 -1.12 -0.62 -2.12 -1.62 -1.05 -0.50 -2.14 -1.63 -1.09 -0.57 2 9 -2.27 -1.71 -1.18 -0.60 -2.32 -1.73 -1.10 -0.53 -2.12 -1.62 -1.05 -0.54 -2.24 -1.69 -1.11 -0.56 3 13 -2.27 -1.70 -1.16 -0.60 -2.18 -1.65 -1.11 -0.58 -2.20 -1.72 -1.09 -0.51 -2.22 -1.69 -1.12 -0.56 4 15 -2.25 -1.69 -1.12 -0.54 -2.19 -1.68 -1.13 -0.60 -2.24 -1.68 -1.14 -0.57 -2.23 -1.68 -1.13 -0.57 5 32 -2.25 -1.68 -1.12 -0.58 -2.16 -1.61 -1.11 -0.60 -2.30 -1.70 -1.15 -0.63 -2.24 -1.66 -1.13 -0.60 6 42 -2.29 -1.76 -1.18 -0.60 -2.26 -1.72 -1.16 -0.62 -2.28 -1.77 -1.16 -0.62 -2.28 -1.75 -1.17 -0.61 7 44 -2.20 -1.66 -1.11 -0.56 -2.26 -1.67 -1.15 -0.57 -2.29 -1.75 -1.16 -0.63 -2.25 -1.69 -1.14 -0.59 8 51 -2.25 -1.70 -1.16 -0.56 -2.27 -1.70 -1.17 -0.62 -2.22 -1.70 -1.13 -0.58 -2.25 -1.70 -1.15 -0.59 9 54 -2.31 -1.77 -1.18 -0.58 -2.19 -1.61 -1.10 -0.58 -2.27 -1.66 -1.15 -0.54 -2.26 -1.68 -1.14 -0.57 10 57 -2.24 -1.66 -1.12 -0.56 -2.20 -1.68 -1.11 -0.55 -2.34 -1.70 -1.19 -0.63 -2.26 -1.68 -1.14 -0.58 11 60 -2.29 -1.74 -1.18 -0.61 -2.16 -1.63 -1.10 -0.54 -2.32 -1.78 -1.20 -0.62 -2.26 -1.72 -1.16 -0.59 12 66 -2.29 -1.76 -1.18 -0.58 -2.23 -1.66 -1.10 -0.52 -2.36 -1.71 -1.14 -0.57 -2.29 -1.71 -1.14 -0.56 13 79 -2.35 -1.76 -1.19 -0.64 -2.21 -1.70 -1.11 -0.54 -2.37 -1.75 -1.17 -0.59 -2.31 -1.74 -1.16 -0.59 14 86 -2.30 -1.74 -1.18 -0.58 -2.25 -1.71 -1.16 -0.58 -2.19 -1.65 -1.09 -0.60 -2.25 -1.70 -1.14 -0.59 15 96 -2.32 -1.71 -1.18 -0.61 -2.35 -1.80 -1.19 -0.60 -2.21 -1.65 -1.13 -0.56 -2.29 -1.72 -1.17 -0.59 表 6 钛同位素标准溶液的定值分析数据
Table 6 The analytical data of Ti isotope composition for reference samples
实验室
编号编号 第1次测定 第2次测定 第3次测定 平均值 δ50Ti δ49Ti δ48Ti δ47Ti δ50Ti δ49Ti δ48Ti δ47Ti δ50Ti δ49Ti δ48Ti δ47Ti δ50Ti δ49Ti δ48Ti δ47Ti ① 2 -2.25 -1.68 -1.12 -0.58 -2.16 -1.61 -1.11 -0.60 -2.30 -1.70 -1.15 -0.63 -2.24 -1.66 -1.13 -0.60 42 -2.29 -1.76 -1.18 -0.6 -2.26 -1.72 -1.16 -0.62 -2.28 -1.77 -1.16 -0.62 -2.28 -1.75 -1.17 -0.61 66 -2.20 -1.68 -1.11 -0.56 -2.26 -1.67 -1.15 -0.57 -2.29 -1.75 -1.16 -0.61 -2.25 -1.70 -1.14 -0.58 ② 15 -2.23 -1.64 -1.15 -0.53 -1.97 -1.51 -1.07 -0.52 -2.29 -1.71 -1.17 -0.56 -2.16 -1.62 -1.13 -0.53 54 -2.05 -1.61 -1.09 -0.51 -2.07 -1.57 -1.11 -0.53 -2.24 -1.63 -1.16 -0.58 -2.12 -1.60 -1.12 -0.54 86 -2.24 -1.70 -1.13 -0.58 -2.08 -1.63 -1.08 -0.55 -2.35 -1.75 -1.21 -0.61 -2.22 -1.70 -1.14 -0.58 ③-1 32 -2.32 -1.75 -1.15 -0.58 -2.44 -1.80 -1.20 -0.63 -2.25 -1.65 -1.14 -0.55 -2.33 -1.73 -1.17 -0.59 44 -2.44 -1.81 -1.22 -0.61 -2.32 -1.71 -1.15 -0.57 -2.41 -1.75 -1.20 -0.61 -2.39 -1.76 -1.19 -0.59 79 -2.47 -1.84 -1.23 -0.64 -2.27 -1.62 -1.10 -0.55 -2.46 -1.81 -1.22 -0.61 -2.40 -1.76 -1.18 -0.60 ③-2 9 -2.13 -1.6 -1.06 -0.54 -1.99 -1.52 -0.99 -0.49 -2.32 -1.76 -1.18 -0.61 -2.15 -1.63 -1.08 -0.55 57 -2.20 -1.63 -1.11 -0.63 -1.94 -1.44 -0.98 -0.52 -2.22 -1.64 -1.09 -0.56 -2.12 -1.57 -1.06 -0.57 96 -2.07 -1.56 -1.05 -0.53 -2.10 -1.56 -1.05 -0.54 -2.26 -1.65 -1.09 -0.53 -2.14 -1.59 -1.06 -0.53 表 7 标准物质定值数据的不确定度
Table 7 Certified values and uncertainty from certified values for reference samples
定值实验室 项目 δ50Ti/‰ δ49Ti/‰ δ48Ti/‰ δ47Ti/‰ 单一实验室 标准值( )-2.26 -1.70 -1.14 -0.58 标准偏差(s) 0.027 0.024 0.017 0.017 不确定度(Uchar) 0.03 0.02 0.024 0.014 测量组数(N) 14 14 14 14 多家实验室 标准值( )-2.23 -1.67 -1.13 -0.57 标准偏差(s) 0.101 0.069 0.045 0.029 不确定度(Uchar) 0.053 0.034 0.02 0.02 测量组数(N) 12 12 12 12 表 8 标准溶液的不确定度
Table 8 The uncertainty of reference samples
不确定度 δ50Ti/‰ δ49Ti/‰ δ48Ti/‰ δ47Ti/‰ Ubb 0.032 0.014 0.011 0.0084 Ults 0.032 0.018 0.013 0.020 Uchar 0.0534 0.0340 0.0239 0.0137 UCRM 0.070 0.041 0.029 0.026 UC 0.140 0.082 0.058 0.051 -
唐索寒,朱祥坤,李津,闫斌.地质样品铜、铁、锌同位素标准物质的研制[J].岩石矿物学杂志,2008,27(4): 279-284. http://www.cnki.com.cn/Article/CJFDTOTAL-YSKW200804003.htm 唐索寒,闫斌,朱祥坤,李津,李世珍.玄武岩标准样品铁铜锌同位素组成[J].岩矿测试,2012,31(2): 218-224. http://www.cnki.com.cn/Article/CJFDTOTAL-YKCS201202003.htm Zhao X M, Zhang H F, Zhu X K, Tang S H, Yan B.Iron isotope evidence for multistage melt-peridotite interactions in the lithospheric mantle of eastern China [J].Chemical Geology,2012,292-293: 127-139. doi: 10.1016/j.chemgeo.2011.11.016
祁昌实,朱祥坤,戴民汉,唐索寒,吴曼,李志红,李世珍,李津.海洋沉积物的铁和锌同位素测定[J].地球化学,2012,41(3): 197-206. http://www.cnki.com.cn/Article/CJFDTOTAL-DQHX201203002.htm Warren P H. Stable-isotopic anomalies and the accre-tionary assemblage of the Earth and Mars: A subordinate role for carbonaceous chondrites [J].Earth and Planetary Science Letters,2011,311: 93-100. doi: 10.1016/j.epsl.2011.08.047
Zhang J J, Dauphas N, Davis A M, Leya I, Fedkin A.The proto-Earth as a significant source of Lunar material [J].Nature Geoscience,2012,5: 251-255. doi: 10.1038/ngeo1429
Zhu X K, Makishima A,Guo Y L, Belshawa N S, O′Nions R K. High precision measurement of titanium isotope ratios by plasma source mass spectrometry [J].International Journal of Mass Spectrometry, 2002,220: 21-29. doi: 10.1016/S1387-3806(02)00767-4
Makishima A, Zhu X K, Belshaw N S, O′Nions R K. Separation of titanium from silicates for isotopic ratio determination using multiple collector ICP-MS [J].Journal of Analytical Atomic Spectrometry, 2002,17: 1290-1294. doi: 10.1039/b204349a
Leya I, Schönbächler M,Wiechert U, KrähenbÜhl U, Halliday A N. High precision titanium isotope measure-ments on geological samples by high resolution MC-ICPMS [J].International Journal of Mass Spectrometry,2007,262: 247-255. doi: 10.1016/j.ijms.2006.12.001
Zhang J J, Dauphas N, Davis A M, Pourmand A.A new method for MC-ICPMS measurement of titanium isotopic composition: Identification of correlated isotope anomalies in meteorites [J].Journal of Analytical Atomic Spectrometry, 2011, 26: 2197-2205. doi: 10.1039/c1ja10181a
丁悌平.稳定同位素测试技术与参考物质研究现状及发展趋势[J].岩矿测试,2002,21(4): 291-300. http://www.cnki.com.cn/Article/CJFDTOTAL-YKCS200204011.htm JJF 1006—94,一级标准物质技术规范[S]. 唐索寒,朱祥坤,赵新苗,李津,闫斌.钛的离子交换分离和MC-ICP-MS高精度同位素组成分析方法[J].分析化学,2011,39(12): 1830-1835. http://cdmd.cnki.com.cn/Article/CDMD-10697-1017028360.htm 全浩, 韩永志.标准物质及其应用技术[M].北京:中国标准出版社,2003. 万德方,李延河,宋鹤彬,刘志坚.硅同位素标准物质的研制[J].地球学报,1997,18(3): 330-335. http://www.cnki.com.cn/Article/CJFDTOTAL-DQXB703.015.htm 杨红梅,段桂玲,凌文黎.La-Ce法岩石标准物质和Ce同位素标准溶液研制[J].地球化学,2009,38(2): 179-186. http://www.cnki.com.cn/Article/CJFDTOTAL-DQHX200902011.htm Craddock P R, Dauphas N.Iron isotopic compositions of geological reference materials and chondrites [J].Geostandards and Geoanalytical Research,2010,35: 101-123.
JJF 1343—2012,标准物质定值的通用原则及统计学原理[S]. 韩永志.标准物质的定值[J].化学分析计量, 2001,10(5): 38-39. http://www.cnki.com.cn/Article/CJFDTOTAL-HOCE201401005.htm 王根荣.标准物质的制备、定值和数据处理[J].上海计量测试, 2002, 29(4): 43-45. http://www.cnki.com.cn/Article/CJFDTOTAL-HNHG805.019.htm 中国实验室国家认可委员会.化学分析中不确定度的评估指南[M].北京:中国计量出版社,2002.
计量