• 中文核心期刊
  • 中国科技核心期刊
  • CSCD来源期刊
  • DOAJ 收录
  • Scopus 收录

微波消解-氢化物发生原子荧光光谱法测定茶园土壤中的铅

Determination of Lead in Tea Garden Soil by Hydride Generation Atomic Fluorescence Spectrometry with Microwave Digestion

  • 摘要: 样品用王水微波消解浸提,氢化物发生-原子荧光光谱法测定茶园土壤中痕量铅的含量。对样品浸取方法、实验条件、增感剂和共存元素进行了条件实验。结果表明,王水微波消解浸取,铅浸出量最大,减少了试剂用量和环境污染;铁氰化钾-盐酸羟胺体系有显著的增感作用,铁氰化钾在配制溶液时用米糠除去试剂中可能存在的铅,降低了空白;钴、锌、砷、镉等共存离子不干扰铅的测定,通过加入邻菲啰啉-硫氰酸钠消除铁和铜的干扰,提高了铅的回收率。方法检出限为0.65 μg/L,精密度(RSD,n=10)为1.89%,回收率在86.8%~110.4%之间。用土壤标准物质验证,测定值与标准值相符,方法快速准确,适合于大批量样品的分析检测。

     

    Abstract: The microwave digestion system with aqua regia was developed for the determination of trace lead in tea garden soil using Hydride Generation-Atomic Fluorescence Spectrometry (HG-AFS). The leaching methods, experimental conditions, booster and coexisting element were optimized. The largest amount of lead was leached out by microwave digestion with aqua regia, which has the advantages of less usage of reagent and less environmental pollution. An appropriate amount of potassium ferricyanide and hydroxylamine hydrochloride improved the hydride generation efficiency of Pb. The lead blank was reduced significantly when Potassium ferricyanide solution was treated with rice husk to remove Pb in the reagent. It was found that Co, Zn, As, and Cd did not interfere with the determination of lead. The addition of 1,10-phenanthroline monohydrate and sodium hydrosulfide could effectively eliminate the interferences from Fe and Cu, which improved the recovery rate of lead. The limit of detection was 0.65 μg/L and the precision was 1.89% (n=10) with recoveries of 86.8%-110.4% for Pb. The reliability of the method has been tested by determination of Pb in the Soils Standard Reference Materials and the results are in agreement with the certified values. The method has been applied to the determination of Pb in tea garden soil with satisfactory results. It was also particularly suitable for large batches of samples.

     

/

返回文章
返回